Thermometrics RTD-11 and RTD-12 immersion probes deliver high-accuracy Pt100 temperature measurement in a direct-immersion format. Thin stainless or Inconel sheath, MgO insulation, and field-replaceable design. Available with PTFE lead wire (RTD-11) or quick-disconnect connector (RTD-12).
| Model | Name | Termination | Process Connection | Best For |
|---|---|---|---|---|
| RTD-11 | Pt100 Immersion Probe — PTFE Lead Wire | 2, 3, or 4-wire PTFE leads, 24 AWG | None (bare probe) | OEM integration, laboratory, DAQ systems |
| RTD-12 | Pt100 Probe — Industrial Connector | 4-pin M12 or standard RTD mini connector | None (bare probe) | Panel instruments, analyzers, quick-disconnect |
| RTD-13 | Pt100 Probe — Tube & Wire, Cut-to-Length | Bare conductors, customer length | None | High-volume OEM; custom sensor harnesses |
| RTD-14 | Pt100 Probe — Threaded Fitting & Lead Wire | PTFE or fibreglass leads | 1/2" NPT (standard) | Process pipes, vessels, general immersion |
| RTD-15 | Replacement Pt100 Element — Spring-Loaded | Bare leads or connector | Hex head for thermowell / industrial housing | Drop-in replacement for RTD-70/RTD-80 assemblies |
| Parameter | Specification |
|---|---|
| Element Type | Pt100 (100 Ω @ 0°C) · Pt1000 on request |
| Temp Coefficient | α = 0.00385 Ω/Ω/°C (IEC 60751) |
| Accuracy Class | Class AA (±0.1°C) · Class A (±0.15°C) · Class B (±0.3°C) |
| Wire Config | 2-wire · 3-wire · 4-wire (Kelvin) |
| Temperature Range | −200°C to 600°C standard; to 850°C with ceramic insulation |
| Sheath OD | 1/8" (3.2mm) · 3/16" (4.8mm) · 1/4" (6.4mm) · custom |
| Sheath Material | 316SS standard · 304SS · Inconel 600 · Hastelloy C276 · Titanium |
| Insulation | MgO (standard) · Ceramic bead for high-temp |
| Lead Wire | PTFE 24 AWG standard · Fibreglass · Ceramic-fiber |
| Response Time (t63) | 1–5 sec typical (sheath OD dependent) |
| Self-Heating Error | <0.1°C at 1 mA excitation (3.2mm OD typical) |
| Calibration | NIST-traceable R vs T table on request |
4-wire (Kelvin) configuration eliminates lead resistance entirely, providing the highest measurement accuracy regardless of cable length. 3-wire compensates for lead resistance assuming balanced leads — suitable for most industrial DCS and PLC applications. Use 4-wire for laboratory, metrology, and any application where lead length exceeds 10 meters.